Single-phase isolating transformers ET1MED** for medical location
Isolating transformers type ET1MED are designed for power supply of medical locations (room group 2), in which there is a danger for health or live of patients in case of passage of a small current through a human body. These transformers meet requirements of standard PN/EN 61558-2-15:2011.
Standard transformers have a reinforced (double) insulation and there is a shield placed between primary and secondary windings connected to an insulated terminal "S". Transformers are provided with two type of sensors for temperature control circuits: posistors type PTC120. Transformers have an additional cental tapping "2M" to connect a relay for control of insulation condition. Upon request, the transformers may be assembled in enclosures with the protection degree of IP 23, IP 44 and IP 54 with the possibility to insert supply cables through glands.
The fixing method of transformers has to ensure an electrical separation from the base plate (for example by means plastic support designed for rated voltage min. 2,5 kV). In order to meet customer expectations we also deliver transofmers together with complete system, which monitors medical network IT
The delivered automatic switching system ensures:
- feeding the reserve supply
- control of basic and reserve voltages
- sygnalling of transformer load
- continuous monitoring of the IT system network insulation resistance
- earth fault localization and signaling
- monitoring of continuity of main circuits of contactor coils
Details:
Design | Single-phase isolating transformers ET1MED are produced in accordance with requirements of PN/EN 61558-2-15:2011 |
---|---|
Insulation class |
T 40E |
Protection degree | IP 00 |
Protection class | II |
Frequency | 50 Hz lub 60 Hz |
Primary voltage | 230 V |
Secondary voltage | 230 V |
Short-circuit voltage | <3% URATED |
No-load current | <3% IRATED |
Leakage current | max. 0,2 mA |
Inrush current | 12 x In max |
Fastening | by means of angles |
Technical sketch
Standard design | |||
Connection diagram of transformer | |||
Dimension table
Technical data of single-phase isolation transformers ET1MED
Type | Power [kVA] |
In [A] |
DELTA °Pfe [W] |
DELTA °Pcu [W] |
uz [%Un] |
io [%In] |
Ibezp [A] gG |
L [mm] |
Bmax [mm] |
H [mm] |
d [mm] |
e [mm] |
k [mm] |
w [mm] |
f [mm] |
Weight [kg] |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
ET1MED - 2,5 | 2,5 | 11,3 | 10 | 86 | 2,72 | 0,95 | 16 | 200 | 180 | 320 | 140 | 145 | 177 | 258 | 11 x 15 | 29 |
ET1MED - 3,15 | 3,15 | 14,1 | 15 | 118 | 2,78 | 0,92 | 25 | 200 | 205 | 320 | 140 | 171 | 203 | 258 | 11 x 15 | 36 |
ET1MED - 4,0 | 4,0 | 18 | 19 | 129 | 2,57 | 0,89 | 25 | 240 | 180 | 365 | 200 | 140 | 178 | 305 | 11 x 16 | 41 |
ET1MED - 5,0 | 5,0 | 22,5 | 22 | 171 | 2,57 | 0,79 | 35 | 240 | 195 | 365 | 200 | 155 | 193 | 305 | 11 x 16 | 46 |
ET1MED - 6,3 | 6,3 | 28,3 | 23 | 212 | 2,42 | 0,67 | 35 | 240 | 195 | 365 | 200 | 155 | 193 | 305 | 11 x 16 | 52 |
ET1MED - 8,0 | 8,0 | 35,7 | 29 | 222 | 2,23 | 0,84 | 50 | 280 | 210 | 420 | 240 | 166 | 208 | 356 | 11 x 15 | 61 |
ET1MED - 10,0 | 10,0 | 45,2 | 30 | 300 | 2,9 | 0,5 | 63 | 280 | 309 | 420 | 240 | 198 | 225 | 356 | 11 x 15 | 80 |
Note:
Manufacturer reserves the right to make changes resulting from the continuous development of products offered.
If requested in advance, it is possible to manufacture transformer in other version.